CVE-2015-3953 | Date: (C)2019-04-06 (M)2023-12-22 |
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVSS Score and Metrics +CVSS Score and Metrics -CVSS V3 Severity: | CVSS V2 Severity: |
CVSS Score : 9.8 | CVSS Score : 10.0 |
Exploit Score: 3.9 | Exploit Score: 10.0 |
Impact Score: 5.9 | Impact Score: 10.0 |
|
CVSS V3 Metrics: | CVSS V2 Metrics: |
Attack Vector: NETWORK | Access Vector: NETWORK |
Attack Complexity: LOW | Access Complexity: LOW |
Privileges Required: NONE | Authentication: NONE |
User Interaction: NONE | Confidentiality: COMPLETE |
Scope: UNCHANGED | Integrity: COMPLETE |
Confidentiality: HIGH | Availability: COMPLETE |
Integrity: HIGH | |
Availability: HIGH | |
| |