CVE-2015-3954Date: (C)2019-04-06   (M)2023-12-22


Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 9.8CVSS Score : 10.0
Exploit Score: 3.9Exploit Score: 10.0
Impact Score: 5.9Impact Score: 10.0
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: NETWORKAccess Vector: NETWORK
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: COMPLETE
Scope: UNCHANGEDIntegrity: COMPLETE
Confidentiality: HIGHAvailability: COMPLETE
Integrity: HIGH 
Availability: HIGH 
  
Reference:
https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01

CWE    1
CWE-285