CVE-2017-12716 | Date: (C)2018-04-26 (M)2023-12-22 |
Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
CVSS Score and Metrics +CVSS Score and Metrics -CVSS V3 Severity: | CVSS V2 Severity: |
CVSS Score : 6.5 | CVSS Score : 3.3 |
Exploit Score: 2.8 | Exploit Score: 6.5 |
Impact Score: 3.6 | Impact Score: 2.9 |
|
CVSS V3 Metrics: | CVSS V2 Metrics: |
Attack Vector: ADJACENT_NETWORK | Access Vector: ADJACENT_NETWORK |
Attack Complexity: LOW | Access Complexity: LOW |
Privileges Required: NONE | Authentication: NONE |
User Interaction: NONE | Confidentiality: PARTIAL |
Scope: UNCHANGED | Integrity: NONE |
Confidentiality: HIGH | Availability: NONE |
Integrity: NONE | |
Availability: NONE | |
| |