[Forgot Password]
Login  Register Subscribe

30479

 
 

423868

 
 

248364

 
 

909

 
 

195388

 
 

282

Paid content will be excluded from the download.


Download | Alert*
CVE
view JSON

CVE-2017-12726Date: (C)2018-02-15   (M)2023-12-22


A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external communications. Smiths Medical assesses that it is not possible to upload files via Telnet and the impact of this vulnerability is limited to the communications module.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 7.3CVSS Score : 7.5
Exploit Score: 3.9Exploit Score: 10.0
Impact Score: 3.4Impact Score: 6.4
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: NETWORKAccess Vector: NETWORK
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: PARTIAL
Confidentiality: LOWAvailability: PARTIAL
Integrity: LOW 
Availability: LOW 
  
Reference:
BID-100665
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A

CWE    1
CWE-798

© SecPod Technologies