CVE-2017-12726 | Date: (C)2018-02-15 (M)2023-12-22 |
A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external communications. Smiths Medical assesses that it is not possible to upload files via Telnet and the impact of this vulnerability is limited to the communications module.
CVSS Score and Metrics +CVSS Score and Metrics -CVSS V3 Severity: | CVSS V2 Severity: |
CVSS Score : 7.3 | CVSS Score : 7.5 |
Exploit Score: 3.9 | Exploit Score: 10.0 |
Impact Score: 3.4 | Impact Score: 6.4 |
|
CVSS V3 Metrics: | CVSS V2 Metrics: |
Attack Vector: NETWORK | Access Vector: NETWORK |
Attack Complexity: LOW | Access Complexity: LOW |
Privileges Required: NONE | Authentication: NONE |
User Interaction: NONE | Confidentiality: PARTIAL |
Scope: UNCHANGED | Integrity: PARTIAL |
Confidentiality: LOW | Availability: PARTIAL |
Integrity: LOW | |
Availability: LOW | |
| |