[Forgot Password]
Login  Register Subscribe

30479

 
 

423868

 
 

248392

 
 

909

 
 

195452

 
 

282

Paid content will be excluded from the download.


Download | Alert*
CVE
view JSON

CVE-2018-10634Date: (C)2018-08-15   (M)2023-12-22


Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G communications between the pump and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 5.3CVSS Score : 2.9
Exploit Score: 1.6Exploit Score: 5.5
Impact Score: 3.6Impact Score: 2.9
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: ADJACENT_NETWORKAccess Vector: ADJACENT_NETWORK
Attack Complexity: HIGHAccess Complexity: MEDIUM
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: NONE
Confidentiality: HIGHAvailability: NONE
Integrity: NONE 
Availability: NONE 
  
Reference:
BID-105044
https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02

CWE    1
CWE-319

© SecPod Technologies