Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 6.5CVSS Score : 3.3
Exploit Score: 2.8Exploit Score: 6.5
Impact Score: 3.6Impact Score: 2.9
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: ADJACENT_NETWORKAccess Vector: ADJACENT_NETWORK
Attack Complexity: LOWAccess Complexity: LOW
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: NONE
Confidentiality: HIGHAvailability: NONE
Integrity: NONE 
Availability: NONE 
  