CVE-2020-12048 | Date: (C)2020-06-30 (M)2023-12-22 |
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
CVSS Score and Metrics +CVSS Score and Metrics -CVSS V3 Severity: | CVSS V2 Severity: |
CVSS Score : 7.5 | CVSS Score : 5.0 |
Exploit Score: 3.9 | Exploit Score: 10.0 |
Impact Score: 3.6 | Impact Score: 2.9 |
|
CVSS V3 Metrics: | CVSS V2 Metrics: |
Attack Vector: NETWORK | Access Vector: NETWORK |
Attack Complexity: LOW | Access Complexity: LOW |
Privileges Required: NONE | Authentication: NONE |
User Interaction: NONE | Confidentiality: PARTIAL |
Scope: UNCHANGED | Integrity: NONE |
Confidentiality: HIGH | Availability: NONE |
Integrity: NONE | |
Availability: NONE | |
| |