[Forgot Password]
Login  Register Subscribe

30389

 
 

423868

 
 

244411

 
 

909

 
 

193363

 
 

277

Paid content will be excluded from the download.


Download | Alert*
CVE
view JSON

CVE-2017-12721Date: (C)2018-02-15   (M)2023-12-22


An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM) attack.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 5.9CVSS Score : 4.3
Exploit Score: 2.2Exploit Score: 8.6
Impact Score: 3.6Impact Score: 2.9
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: NETWORKAccess Vector: NETWORK
Attack Complexity: HIGHAccess Complexity: MEDIUM
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: NONE
Confidentiality: HIGHAvailability: NONE
Integrity: NONE 
Availability: NONE 
  
Reference:
BID-100665
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A

CWE    1
CWE-295

© SecPod Technologies