CVE-2017-12725Date: (C)2018-02-15   (M)2023-12-22


A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 5.6CVSS Score : 6.8
Exploit Score: 2.2Exploit Score: 8.6
Impact Score: 3.4Impact Score: 6.4
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: NETWORKAccess Vector: NETWORK
Attack Complexity: HIGHAccess Complexity: MEDIUM
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: PARTIAL
Confidentiality: LOWAvailability: PARTIAL
Integrity: LOW 
Availability: LOW 
  
Reference:
BID-100665
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A

CWE    1
CWE-798