CVE-2020-6964 | Date: (C)2020-01-25 (M)2023-12-22 |
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X and CARESCAPE Central Station (CSCS) Versions 2.X, the integrated service for keyboard switching of the affected devices could allow attackers to obtain remote keyboard input access without authentication over the network.
CVSS Score and Metrics +CVSS Score and Metrics -CVSS V3 Severity: | CVSS V2 Severity: |
CVSS Score : 8.6 | CVSS Score : 5.0 |
Exploit Score: 3.9 | Exploit Score: 10.0 |
Impact Score: 4.0 | Impact Score: 2.9 |
|
CVSS V3 Metrics: | CVSS V2 Metrics: |
Attack Vector: NETWORK | Access Vector: NETWORK |
Attack Complexity: LOW | Access Complexity: LOW |
Privileges Required: NONE | Authentication: NONE |
User Interaction: NONE | Confidentiality: NONE |
Scope: CHANGED | Integrity: PARTIAL |
Confidentiality: NONE | Availability: NONE |
Integrity: HIGH | |
Availability: NONE | |
| |