[Forgot Password]
Login  Register Subscribe

30480

 
 

423868

 
 

251625

 
 

909

 
 

196370

 
 

282

Paid content will be excluded from the download.


Download | Alert*
CVE
view JSON

CVE-2017-12718Date: (C)2018-02-15   (M)2023-12-22


A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.

CVSS Score and Metrics +CVSS Score and Metrics -

CVSS V3 Severity:CVSS V2 Severity:
CVSS Score : 8.1CVSS Score : 6.8
Exploit Score: 2.2Exploit Score: 8.6
Impact Score: 5.9Impact Score: 6.4
 
CVSS V3 Metrics:CVSS V2 Metrics:
Attack Vector: NETWORKAccess Vector: NETWORK
Attack Complexity: HIGHAccess Complexity: MEDIUM
Privileges Required: NONEAuthentication: NONE
User Interaction: NONEConfidentiality: PARTIAL
Scope: UNCHANGEDIntegrity: PARTIAL
Confidentiality: HIGHAvailability: PARTIAL
Integrity: HIGH 
Availability: HIGH 
  
Reference:
BID-100665
BID-101252
EXPLOIT-DB-43776
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A

CWE    1
CWE-119

© SecPod Technologies